Digital Manometer MT300 | Yokogawa Test&Measurement

High precision measurements

High resolution display (When /R1 is selected.)

By using a high resolution display, pressure measurements can be more accurately displayed and visualized. In addition, the increased resolution, especially in calibration environments, reduces the uncertainty of the entire calibration process.

High resolution display

High speed measurement (When /F1 is selected.)

Rapid pressure changes and transient responses requires high speed and reliable measurements. Depending on the application, you can choose your measurement speeds from three different modes, normal, medium or high speed.

Synchronous measurement

Using the synchronization features, you can collect data and display measurements from multiple units. Measurements can be captured and correlated at high speed with high accuracy and confidence.

Application:
In the performance testing of vacuum cleaners, the suctionforce is calculated from the amount of suction-air in the pressure equalizing container and the degree of vacuum. The amount of intake-air can be obtained by measuring the differential pressure via a pitot tube, and the degree of vacuum can be obtained by measuring the absolute pressure. With the /F1 option, the dynamic characteristics can be captured at high speed, with accurate results, by synchronously measuring these two values from two MT300s.

Synchronous measurement

Support for efficient works

Leak test feature

The Leak-test feature can be used to measure pressure change or leak rate within the measurement period. It can be used to check the tightness or integrity of a pressure measurement system.

Leak test

Start: Pressure value and time when started
Stop: Pressure value and time when stopped
Time: Measurement time
Delta: Difference of pressure value between started and stopped
Rate: Deffence of pressure value per minute

Leak test

Scaling

The scaling function can be used to assign customer specific coefficients to the measured values. Depending on the application, you can display your own conversion value. Scaling coefficient A and offset B are determined according to the following equation. Set the scale values for the upper and lower span limits.

Scaling

※ If you change the unit, set scaling coefficient A and offset B again.

Statistical processing

You can apply statistical processing to the data acquired. Find and display the maximum value, minimum value, average value, and standard deviation for the measured data. When error data is detected, the number of error data within the measurement range can be recorded and displayed.

統計処理

Max:Maximum value
AVG:Average value
ERR:Number of error data

Min:Minimum value
σ:Standard deviation
 

Support for linkage with external devices

D/A output (When /DA is selected)

The Digital-to-Analog (D/A) option, makes it easy to output measured pressure values to an external terminal going to a measurement system or a recorder. The output update interval can be set to 250 μs in combination with the /F1 option (in medium-speed/high-speed modes).

Scaling

Example of the waveform differences in measurement modes

Comparator output

Using external I/O terminals, you can output control signals, based on set upper and lower limits and judgement criteria’s on measured values. These features allow automation of production and/or inspection lines for pressure-related products.

Interfaces

Communication Interfaces such as GPIB, USB (type-B), and ETHERNET are available as standard features. Communication commands are compatible*1, *2 with existing models*3, making it easy to expand or update your existing measurement system.
*1: Some command cannot be used.
*2: Compatible commands cannot be used when -G07, -G08, -A05, -A06 are selected.
*3: MT210/MT210F/MT220

Support for field device calibration and maintenance works

Calibration involves inputting the same pressure level to both a calibrator and a transmitter and comparing the transmitter output with a value measured by the calibrator. The MT300 come with the functions you need for such calibration or maintenance work in the field.

  • Outputting 24 VDC for the supply of transmitter
  • Measuring transmitter output (1 to 5 V or 4 to 20 mA selective)
  • Built-in communication resistance enables ON/OFF switching.

Above functions can be available when /DM is selected.

  • It is possible to bring it out without AC power by Li-ion battery operation.
    Running time: Approx. 6 hours with all functions turned on Charge time: Approx. 6 hours

Above function can be available when /EB is selected or add them as accessory after ordered.